Features
The Unfors Mult-O-Meter offers a cost effective and rapid way for measurements. The meter is designed with the aim to make the measurement procedure simple and straightforward. The Unfors Mult-O-Meter is extremely easy to use. Only two buttons are used to control the instrument: On/Off and Parameter for scrolling the measured values. No set-up time is needed. Just position the detector in the radiation field and expose. The specially designed LCD clearly indicates measured value and unit together.
External Detector: kVp DENT, Dose DENT, Rate DENT, Time, Pulse and Extended Functions
Specifications
kVp DENT Range (auto): 45 – 110 kVp
Minimum exposure: 3 mA at 70 kVp, 50 cm
Dose DENT Range (auto): 8 µGy- 9999 Gy, (1 mR – 9999 R)
Energy dependence: < 5 % at 2.5 mm Al, (50 – 150 kVp)
Rate DENT Range (auto): 8 µGy/s – 80 mGy/s, (55 mR/min – 550 R/min)
Energy dependence: < 5 % at 2.5 mm Al, (50 – 150 kVp)
Time Range (auto): 1 ms – 9999 s measured with the trigger detector
Inaccuracy: 0.5 % or 3 digits
Pulse Range: 1 ms – 9999 s measured with the trigger detector
Inaccuracy: 1 pulse
Parts
Extended functions:
Normalization of dose,
Dose reproducibility,
mAs linearity,
Simplified HVL ,
Time corresponding to the 75 % triglevel of kVp waveform
kVp with a 5, 50, 150 ms delay
Case Unfors QA Case 1
Image Gallery