Features
The Unfors Mult-O-Meter offers a cost effective and rapid way for measurements. The meter is designed with the aim to make the measurement procedure simple and straightforward. The Unfors Mult-O-Meter is extremely easy to use. Only two buttons are used to control the instrument: On/Off and Parameter for scrolling the measured values. No set-up time is needed. Just position the detector in the radiation field and expose. The specially designed LCD clearly indicates measured value and unit together.
Internal Detector: kVp MAM, Dose MAM, Rate MAM and Time
Specifications
kVp MAM Range (auto): 22 – 40 kVp
Minimum exposure: 20 mA at 28 kVp, 65 cm
Dose MAM Range (auto): 100 µGy – 9999 Gy, (10 mR – 9999 R)
Energy dependence: 5 % at 30 µm Mo, (25 – 30 kVp)
Rate MAM Range (auto): 0.1 mGy/s – 500 mGy/s, (700 mR/min – 3500 R/min)
Energy dependence: &plusm; 5 % at 30 m Mo, (25 – 30 kVp)
Time Range (auto): 1 ms – 9999 s measured with the trigger detector
Inaccuracy: 0.5 % or 3 digits
Parts
Case Unfors QA Case 1