Unfors MaM Detector

Detachable multi-parameter detector with Active Compensation.

Unfors MaM Detector
 
  • Features
  • Specifications
  • Image Gallery

Features

Detachable multi-parameter detector with Active Compensation. Mo/Mo, Mo/Rh, Mo/Al, Rh/Rh, Rh/Al and W/Rh beam qualities incl. Certificates.

Specifications

Unfors Xi Mammography Detector

Size: 12 x 22 x 117 mm (0.5 x 0.9 x 4.6 in)

Weight: 50 g (2 oz)

Dose

Beam qualities: Basic: Mo/Mo, Mo/Al, Mo/Rh, Rh/Rh, Rh/Al, W/Rh

M-Pro option: W/Rh, Mo/Rh, W/Ag

Scanning mammo option: W/Al

Range: 5 μGy – 9999 Gy (0.5 mR – 9999 R)

Trig level: 10 μGy/s (70 mR/min)

Uncertainty: 5 % (22 – 40 kV (Mo/Mo: 20 – 40kV), 0 – 2.5 mm Al added filtration,

Active Compensation) or &plusm; 5 μGy (0.5 mR)

Uncertainty (W/Al): 5 % (22 – 40 kV, 0.5 mm Al total filtration) or &plusm; 5 μGy (0.5 mR)

Dose rate

Range: 10 μGy/s – 100 mGy/s (70 mR/min – 700 R/min)

Trig level: 10 μGy/s (70 mR/min)

Uncertainty: 5 % (22 – 40 kV (Mo/Mo: 20 – 40kV) and for HVL up to 2.5 mm Al

added filtration, Active Compensation) &plusm; 5 μGy/s (35 mR/min)

Uncertainty (W/Al): 5 % (22 – 40 kV + 0.5 mm Al total filtration) or &plusm; 5 μGy/s (35 mR/min)

kV

Beam qualities: Basic: Mo/Mo

M-Pro option: Mo/Rh, W/Rh

Scanning mammo option: W/Al

Range (W/Rh): 23 – 40 kV

Uncertainty (W/Rh): 2 % or 0.7 kV (Active Compensation for inherent Rh filtration

of 55 – 60 μm. User selectable paddle compensation.)

Range: 20 – 40 kV

Uncertainty (Mo/Mo): 2 % or 0.7 kV (Active Compensation for inherent Mo filtration

of 25 – 35 μm. User selectable paddle compensation.)

Uncertainty (Mo/Rh): 2 % or 0.7 kV (Active Compensation for inherent Rh filtration

of 25 – 30 μm. )

Uncertainty (W/Al): 2 % or 0.7 kV (0.5 mm Al total filtration)

Sensitivity: 10 mA, 28 kV, 65 cm (25.6 in), no added filtration

Exposure time

Range: 1 ms – 999 s

Uncertainty: 0.5 % or 0.2 ms

HVL

Beam qualities: Basic: Mo/Mo, Mo/Al, Mo/Rh, Rh/Rh, Rh/Al, W/Rh

M-Pro option: W/Rh, Mo/Rh, W/Ag

Scanning mammo option: W/Al

Range: 0.2 – 1.2 mm Al, depending on beam quality

Uncertainty: 5 % (for up to 2.5 mm Al added filtration to each beam quality)

Range (W/Al): 0.32 – 0.58 mm Al

Uncertainty (W/Al): 5 % (with 0.5 mm Al total filtration)

Waveform

Bandwidth: 0.3 kHz

Memory depth: 1200 ms

Image Gallery

Click on a thumbnail below to view full size.

  • Default Image

Posted in Unfors | Tagged assurance, control, detection, diagnostic, dose, dose-frame, dose-rate, hvl, kvp, mammo, mammography, pulse, pulse-rate, qa, qc, quality, time-and-waveforms, Unfors, x-ray, xi, xray

Next »
  • Manufacturers
  • Technologies

Featured Manufacturers

  • Biodex
  • Bionix
  • Capintec
  • Diacor
  • Exini
  • Fluke
  • International Specialty Products
  • IZI Medical Products
  • Mediso
  • PTW
  • Topex
  • Unfors
  • Vision RT
  • View all...

Technologies

  • Nuclear Medicine
  • Radiology
  • Radiation Therapy

Copyright © 2012 Harpell Associates Inc..

  • Home
  • About
  • News
  • Contact