Features
Unfors R/F X-ray System 40 is a dedicated System when you need a small and truly portable solution for service work on fluoroscopic machines and only a sensitive dose range is needed. The Unfors Mult-O-Meter 514L measures kVp, dose, rate, dose/frame and time simultaneously with the most sensitive detectors Unfors offer. By means of the Dose/Frame feature it is possible to easily check machines with pulsed fluoroscopy capabilities.
Unfors Mult-O-Meter 514L with external detector for kVp FLUORO, Dose FLUORO, Rate FLUORO, Time, Dose/Frame, IR Link and Extended Functions
Specifications
kVp FLUORO Range (auto): 45 – 155 kVp
Minimum exposure: 1.5 mA at 70 kVp, 50 cm
Dose FLUORO Range (auto): 100 nGy – 9999 Gy, (11 µR – 9999 R)
Energy dependence: < 5 % at 2.5 mm Al, (50 – 150 kVp)
Rate FLUORO Range (auto): 100 nGy/s – 0.5 mGy/s, (0.7 mR/min – 3.5 R/min)
Energy dependence: < 5 % at 2.5 mm Al, (50 – 150 kVp)
Time Range (auto): 1 ms – 9999 s measured with the trigger detector
Inaccuracy: 0.5 % or 3 digits
Dose/Frame Range (auto): 200 nGy – 9999 Gy, (23 mR – 9999 R)
Minimum pulses/s: 10
Minimum pulse width: 1 ms
Extended functions Normalization of dose
Dose reproducibility
mAs linearity
Simplified HVL
Time corresponding to the 75 % triglevel of kVp waveform
kVp with a 5, 50, 150 ms delay
IR Link
Equipped with an infrared interface for transmission of measured data to a
PC. The IR receiver connected to the serial port of a PC, receives the
measured data. For data collection and calculations, the Unfors Excel Add-in
software is available for user customized reports.
Unfors DAQ-10 Software – needs.
Operating system Windows 98/ME, Windows 2000, Windows XP.
Microsoft Excel Office 97, 2000 and XP
Pen detector
holder Extendable holder for Unfors Mult-O-Meter external detectors for
easy positioning on I.I.’s, in buckys etc.
Options
1503010 Unfors Mult-O-Meter 704 SO (mA, mAs, Time)
1503011 Unfors Mult-O-Meter 714 SO (mAs)
1503012 Unfors Mult-O-Meter 603 SO (Dose CT, Time)
1901043 HVL Kit
Parts
Unfors Mult-O-Meter 514L
Ext. detector 1: kVp FLUORO, Dose FLUORO, Rate FLUORO, Time, incl. EF, IR link and Dose/Frame
Unfors DAQ-10 Kit
Pen detector holder
Unfors QA Case 2
Image Gallery
Click on a thumbnail below to view full size.

Comments are closed.