Features
- Two in one – Combines a sensitometer and a manual densitometer in one compact unit
- Exposes a 21-step wedge on blue or green sensitive X-ray films
- Suitable for manual single point measurement of optical densities in the range (0.00 … 4.5) OD
The SensodensiX device is a combination of a sensitometer and a manual densitometer for quality control of X-ray film processing and for checking test phantom films.
The single-sided, dual color, light-emitting sensitometer exposes a step wedge of 21 steps on X-ray films. The light emission can be selected for blue sensitive films (spectral maximum 460 nm) and green sensitive films (spectral maximum 510 nm). The wedge constant is 0.15 OD according to international standards.
The manually operated densitometer is designed for optical density single point measurements of processed X-ray films, exposed by a sensitometer or by X-ray equipment. The measuring accuracy and the repeatability are ± 0.01 OD each. The measuring aperture is 3 mm diameter. The length of the measuring arm is 20 cm, which makes it possible to measure the optical density even in the middle of large size films (35 cm x 43 cm).
An external power supply is required to operate the unit.
Parts
L981266 SensodensiX combined unit
Image Gallery