Hp(10) Secondary Standard Chamber Type 34047

Parallel plate ionization chamber for direct measurement of Hp(10) personal dose equivalent on a slab phantom

Hp(10) Secondary Standard Chamber Type 34047
 

Features

  • Vented sensitive volume of 10 cm3
  • Measures the Hp(10) personal dose equivalent directly
  • Suitable as a reference chamber for Hp(10) calibration

The parallel plate ionization chamber model 340351 is integrated into a slab phantom to measure the Hp(10) radiation protection measuring quantity directly. The high performance chamber is designed to be used as a secondary standard chamber for calibration purposes. The beam calibration with the Hp(10) chamber makes it unnecessary to precisely determine the spectrum of the X-ray beam. The chamber comes uncalibrated; a primary standard calibration by PTB, the German National Laboratory, is available. The chamber set includes a phantom slab of 31 mm thickness with chamber assembly and an additional PMMA phantom slab of 120 mm thickness. Both sets available include an adapter cable to connect the chamber either to a dosemeter with M connector or with BNC connector and banana pin. The Hp(10) chamber should be used in connection with a high quality dosemeter such as UNIDOS, UNIDOS E or UNIDOSwebline to ensure best performance.

Specifications

Type of product vented parallel plate chamber
Application radiation protection measurements
Measuring quantity Hp(10) personal dose equivalent
Nominal sensitive volume 10 cm3
Design not waterproof, vented
Reference conditions 20°C, 1013 hPa 65 % rel. humidity
Reference point chamber center, 13.5 mm below chamber surface or 15.5 mm below surface of integrated step cylinder
Nominal response

350 nC/Sv

Chamber voltage 400 V nominal
Leakage current

≤ ± 10 fA

Cable leakage ≤ 1 pC/(Gy·cm)

Materials and measures

Phantom material PMMA
Outer dimensions 300 mm x 300 mm

chamber assembly

height 31 mm

additional slab phantom height 120 mm

Useful Ranges

Chamber voltage ± (300 … 500) V
Radiation quality (15 … 1400) keV

Temperature (10 … 40) °C
(50 … 104) °F
Humidity

(10 … 80) %, max 20 g/m3

Air pressure (700 … 1060) hPa

Options

PTB Primary standard calibration upon request

  1. Ankerhold, Ambrosi, Eberle ? A chamber for determining the conventionally true value of Hp(10) and H*(10) needed by calibration laboratories ? Rad. Prot. Dos. Vol. 96, Nos 1-3, pp. 133 – 137 (2001), Nucl. Techn. Publishing

Parts

L981937 Hp(10) Secondary standard chamber type 34035, connecting system M

L981938 Hp(10) Secondary standard chamber type 34035, connecting system BNC and banana pin

Image Gallery

Click on a thumbnail below to view full size.

  • Default Image