Features
- Makes it possible to measure doses of all types of dental, intraoral, remote and panoramic X-ray units
- Provides all DIADOS features by operating in conjunction with the DIADOS dosemeter
- Calibration factors for dental use are stored in the DIADOS data base
The different types of dental X-ray units require different setup for dose measurements. Transmission doses from intraoral units are determined behind a 6 mm Al absorber. Remote X-ray units require the dose measurement behind an absorber of 6 mm Al and 0.8 mm Cu. Alternative filters of 0.3 mm Cu and 8 mm PTFE can be used. Dose measurements of panoramic X-ray units with movable film cassette are made by means of a cassette adapter of 15 cm x 30 cm size to which the detector is mounted. The filter setting for this type of dose measurement is 6 mm Al + 0.5 mm Cu for film-screen systems and 0.8 mm Cu for digital systems. All filters required are included in the absorber set T42018. The filters are placed between focus and detector. Dental doses are measured using the DIADOS diagnostic detector connected to the DIADOS dosemeter. The calibration for dental use has to be ordered separately. All appropriate calibration factors are stored in the DIADOS data base.
Parts
T60004 DIADOS diagnostic detector
E21232 Diagnostic dental detector calibration
T42018 Absorber set for dental dose measurement
T20003 Dental cassette adapter
L981234 DIAset DENT