Features
- Makes CTDI1 and CTDIW2 determination possible according to IEC 61223-2-6
- CT ion chamber, sensitive length 10 cm
- Body phantom, 32 cm ∅, and head phantom, 16 cm ∅, both acrylic cylinders of 15 cm height
- Combined head and body phantom is available
The CT probe is inserted for measurements into one of five holes of the head or body phantom, which represent the body part to be CT scanned. Holes not used are filled by acrylic dummy plugs, and a support keeps the phantom in its position on the CT couch. Etched crosshairs on the phantoms allow exact alignment. The CT chamber can be supplied by a BNT or TNC or PTW-M connector for operation in connection with a UNIDOS E dosemeter. Only the DIADOS and DIADOS E require an optional HV adapter. The measuring quantity is dose length product in Gy·m.
- Computed Tomography Dosimetry Index;
- Weighted CTDI
Options
T16018 DIADOS CT adapter for chamber HV
T40016.1.010 Carrying case for body phantom
T40017.1.010 Carrying case for head phantom, DIADOS and CT chamber
T40027.1.050 Carrying case for head and body phantom, DIADOS E and CT chamber
Parts
CT chamber, connecting system BNT, TNC or M:
30009 CT chamber
T40016 CT body measuring phantom
T40017 CT head measuring phantom
T40027 CT head and body phantom
Image Gallery
Click on a thumbnail below to view full size.