75 Cm3 SFD Diagnostic Chamber Type 34060

Shadow-free plane parallel chamber for absolute dosimetry in diagnostic radiology

75 Cm3 SFD Diagnostic Chamber Type 34060
 

Features

  • Shadow-free design for use with automatic exposure control / brightness control
  • Sensitive volume 75 cm3, vented to air
  • Suitable for measurements in front of and behind a phantom

The 34060 SFD diagnostic chamber is a general purpose, high precision chamber for measurements in diagnostic radiology. The chamber complies with the standard IEC 61674. Its shadow-free design makes it possible to use the chamber even while the automatic exposure control or brightness control is activated. Together with an adequate diagnostic dosemeter the chamber features a wide dynamic range for measurements either in front of or behind a patient-equivalent phantom. The length of the mounted connection cable is 2.5 m.

Specifications

Type of product vented plane parallel ionization chamber acc. IEC 61674
Application absolute dosimetry in diagnostic radiology
Measuring quantities air kerma, exposure
Reference radiation quality 70 kV, HVL 2.58 mm Al (RQR5)
Nominal sensitive volume 75 cm3
Design not waterproof, vented

Reference point in chamber center
Direction of incidence perpendicular to chamber plane, see label ‘Focus’

Nominal response 2.8 µC/Gy
Long-term stability ≤ 1 % per year
Chamber voltage 200 V nominal
± 400 V maximal

Polarity effect at RQR/RQA5 ≤ 1 %

Directional response ≤ 3 % for chamber tilting up to ≤ ± 15°
Leakage current ≤ ± 5 fA

Cable leakage ≤ 1 pC/(Gy·cm)

Materials and measures

Entrance window 0.52 mm PMMA,
1.19 g/cm3
0.002 mm graphite
0.32 g/cm3

Total window area density 62 mg/cm2
Water-equivalent window thickness 0.6 mm

Sensitive volume radius 45.7 mm
depth 2 x 5.71mm

Ion collection efficiency at nominal voltage

Ion collection time 1 ms
Max. dose rate for
≥ 99 % saturation
≥ 95 % saturation

0.16 Gy/s
0.78 Gy/s

Useful Ranges

Chamber voltage ± (100 … 400) V
Radiation quality (25 … 150) kV X-rays
Field size (11 x 11) cm2 … (40 x 40) cm2

Temperature (10 … 40) °C
(50 … 104) °F
Humidity (10 … 80) %, max 20 g/m3

Air pressure (700 … 1060) hPa

Parts

TN34060-2,5 SFD diagnostic chamber 75 cm3, connecting system BNT

TW34060-2,5 SFD diagnostic chamber 75 cm3, connecting system TNC

TM34060-2,5 SFD diagnostic chamber 75 cm3, connecting system M

Image Gallery

Click on a thumbnail below to view full size.

  • Default Image