30 Cm3 Cylinder Stem Ionization Chamber Type 23361

Cylindrical PMMA ionization chamber with a long rigid stem for radiation protection measurement

30 Cm3 Cylinder Stem Ionization Chamber Type 23361
 

Features

  • Vented sensitive volume of 30 cm3
  • Suitable as high precision reference chamber for radiation protection dosimetry
  • Very flat energy response within a wide range
  • Radioactive check device (option)

The cylinder stem chamber is designed as a reference chamber for absolute dosimetry to be used by secondary standard dosimetry laboratories (SSDL) and users with high quality requirements. It has very small variations of response with radiation quality from low X-ray energies up to high-energy photon radiation. The guard ring is designed up to the sensitive volume. The chamber is constructed with a long rigid stem of approx. 20 cm length for easy mounting in the radiation beam. An acrylic build-up cap with 3 mm wall thickness for in-air measurement in 60Co beams is included with each chamber, as well as a calibration certificate. Air density correction is required for each measurement. A radioactive check device and an appropriate holding device are available.

Specifications

Type of product vented cylindrical ionization chamber
Application radiation protection measurements
Measuring quantity photon equivalent dose, exposure

Nominal sensitive volume 30 cm3
Design

not waterproof, vented

Reference point on chamber axis,
27 mm from chamber tip
Direction of incidence radial

Nominal response 1 µC/Gy
Chamber voltage 400 V nominal
± 500 V maximal
Energy response < ± 4 % (40 keV … 60Co)
Directional response in air ≤ ± 0.5 % for for rotation around the chamber axis for tilting see diagram page 63

Leakage current ≤ ± 10 fA

Stem leakage ≤ 1 pC/(Gy·cm)

Materials and measures

Wall of sensitive volume 1 mm PMMA, graphited
Total wall area density 119 mg/cm2
Dimension of sensitive volume radius 15.5 mm
length 51 mm
Central electrode graphite coated Al, diameter 14 mm
Outer dimensions diameter 33 mm
length 335 mm
Build-up cap PMMA, thickness 3 mm

Ion collection efficiency at nominal voltage

Ion collection time 1.3 ms
Max. dose rate for
≥ 99.5 % saturation
≥ 99.0 % saturation

60 mGy/s
120 mGy/s

Max. dose per pulse for
≥ 99.5 % saturation
≥ 99.0 % saturation

50 µGy
100 µGy

Useful Ranges

Radiation quality 30 keV … 1.3 MeV
Temperature (10 … 40) °C
(50 … 104) °F
Humidity (10 … 80) %, max 20 g/m3

Air pressure (700 … 1060) hPa

Options

T48010 Radioactive check device 90Sr

T23237 Chamber holding device for check device

Parts

TN23361 Cylinder stem chamber 30 cm3,connecting system BNT

TW23361 Cylinder stem chamber 30 cm3, connecting system TNC

TM23361 Cylinder stem chamber 30 cm3, connecting system M