3.14 Cm3 CT Chamber Type 30009

Vented cylindrical pencil chamber for dose length product measurements in computed tomography

3.14 Cm3 CT Chamber Type 30009
 

Features

  • Pencil type chamber for measurements within a CT head or body phantom or in air
  • Provides a sensitive measuring length of 10 cm
  • Shows a homogeneous response over the whole chamber length

The CT chamber type 30009 is a vented cylinder chamber designed for measurements of photon radiation in computed tomography. The typical measuring quantity is dose length product, which allows the determination of the CTDI1 and CTDIW2 according to IEC 61223-2-6

Specifications

Type of product vented pencil type chamber
Application dosimetry in computed tomography
Measuring quantities air kerma length product, exposure length product
Reference radiation quality 120 kV, HVL 5.00 mm Al (RQR9)
Nominal sensitive volume

3.14 cm3

Design not waterproof, vented, pencil type
Reference point chamber center

Direction of incidence radial
Nominal response 14 nC/Gy
Chamber voltage 100 V nominal
± 500 V maximal
high voltage to be connected only with active current-limiting device (Imax < 0.5 mA)
Energy response ≤ ± 3 % (70 … 150) kV
Directional response ≤ ± 2 % for tilting < ± 170°

Leakage current ≤ ± 10 fA
Cable leakage ≤ 1 pC/(Gy·cm)

Materials and measures

Wall material 1 mm PMMA, graphite coated
Wall area density 120 mg/cm2
Dimension of sensitive volume radius 7 mm
length 100 mm
Electrode Al tube, outer diameter 3 mm, inner diameter 2 mm

Ion collection efficiency at nominal voltage

Ion collection time

300 µs

Max. dose rate for
≥ 99.5 % saturation
≥ 99.0 % saturation

0.9 Gy/s
1.9 Gy/s

Max. dose per pulse for
≥ 99.5 % saturation
≥ 99.0 % saturation

0.14 mGy
0.35 mGy

Useful Ranges

Chamber voltage ± (100 … 400) V
Radiation quality (70 … 150) kV

Temperature (10 … 40) °C
(50 … 104) °F
Humidity (10 … 80) %, max 20 g/m3

Air pressure (700 … 1060) hPa

Parts

TN30009 CT chamber 3.14 cm3, connecting system BNT

TW30009 CT chamber 3.14 cm3, connecting system TNC

TM30009 CT chamber 3.14 cm3, connecting system M

  1. CTDI = Computed Tomography Dose Index
  2. CTDIW = Weighted Computed Tomography Dose Index
  • Both definitions acc. IEC 60601-2-44