1.0 Cm3 Rigid Stem Chamber Type 23331

High-volume therapy chamber for absolute dosimetry in high-energy photon and electron beams

1.0 Cm3 Rigid Stem Chamber Type 23331
 

Features

  • Fully guarded chamber
  • Sensitive volume 1.0 cm3, vented to air
  • Acrylic wall, graphited Aluminum central electrode
  • Radioactive check device (option)

The 23331 rigid stem chamber is used for absolute dose measurements in radiation therapy. Correction factors needed to determine absorbed dose to water or air kerma are published in the pertinent dosimetry protocols. The acrylic chamber wall ensures the ruggedness of the chamber. The chamber is designed for the use in solid state phantoms and is therefore not waterproof.

Specifications

Type of product vented cylindrical ionization chamber
Application absolute dosimetry in radiotherapy beams
Measuring quantities absorbed dose to water, air kerma, exposure

Reference radiation quality 60Co
Nominal sensitive volume 1.0 cm3

Design not waterproof , vented, fully guarded
Reference point on chamber axis, 11.5 mm from chamber tip
Direction of incidence radial
Nominal response 33 nC/Gy
Long-term stability ≤ 1 % per year
Chamber voltage 400 V nominal
± 500 V maximal
Polarity effect at 60Co < 0.5 %
Photon energy response < ± 2 % (70 kV … 250 kV)
< ± 4 % (200 kV … 60Co)

Directional response in solid state phantom < ± 0.5 % for rotation around the chamber axis, < ± 1 % for tilting of the axis up to ± 30°
Leakage current ≤ ± 4 fA
Cable leakage ≤ 1 pC/(Gy·cm)

Materials and measures

Wall of sensitive volume 0.4 mm PMMA,
1.19 g/cm3

0.15 mm graphite,
0.82 g/cm3

Total wall area density 59.9 mg/cm2
Dimension of sensitive volume radius 7.9 mm
length 22 mm

Central electrode Al 99, diameter 1.5 mm
Build-up cap

PMMA, thickness 3 mm

Ion collection efficiency at nominal voltage

Ion collection time 200 µs
Max. dose rate for
≥ 99.5 % saturation
≥ 99.0 % saturation
1.8 Gy/s
3.7 Gy/s
Max. dose per pulse for
≥ 99.5 % saturation
≥ 99.0 % saturation
190 µGy
460 µGy

Useful Ranges

Chamber voltage ± (100 … 400) V
Radiation quality 70 keV … 50 MV photons
(10 … 45) MeV electrons
Field size (5 x 5) cm2 … (40 x 40) cm2
Temperature (10 … 40) °C
(50 … 104) °F
Humidity (20 … 75) %, max 20 g/m3

Air pressure (930 … 1050) hPa

Options

T48012 Radioactive check device 90Sr

T48002.3.004 Chamber holding device for check device

Parts

TN23331 Rigid stem chamber 1.0 cm3, connecting system BNT

TW23331 Rigid stem chamber 1.0 cm3, connecting system TNC

TM23331 Rigid stem chamber 1.0 cm3, connecting system M

Image Gallery

Click on a thumbnail below to view full size.

  • Default Image