0.3 Cm3 Rigid Stem Chamber Type 23332

Therapy chamber for absolute dosimetry in high-energy photon and electron beams

0.3 Cm3 Rigid Stem Chamber Type 23332
 

Features

  • Fully guarded chamber
  • Sensitive volume 0.3 cm3, vented to air
  • Acrylic wall, graphited
  • Aluminum central electrode
  • Radioactive check device (option)

The 23332 chamber is used for absolute dose measurements in radiation therapy in cases where the high volume of the 23331 chamber is not needed and a higher spatial resolution is needed. Correction factors needed to determine absorbed dose to water or air kerma are published in the pertinent dosimetry protocols. The acrylic chamber wall ensures the ruggedness of the chamber. The chamber is designed for the use in solid state phantoms and is therefore not waterproof.

Specifications

Type of product vented cylindrical ionization chamber
Application absolute dosimetry in radiotherapy beams
Measuring quantities

absorbed dose to water, air kerma, exposure

Reference radiation quality 60Co
Nominal sensitive volume

0.3 cm3

Design not waterproof, vented, fully guarded
Reference point

on chamber axis, 9.5 mm from chamber tip

Direction of incidence radial

Nominal response 10 nC/Gy
Long-term stability

≤ 1 % per year

Chamber voltage 400 V nominal
± 500 V maximal
Polarity effect at 60Co < 0.5 %
Photon energy response < ± 2 % (140 kV … 250 kV)

< ± 4 % (200 kV … 60Co)

Directional response in solid state phantom < ± 0.5 % for rotation around the chamber axis,

< ± 1 % for tilting of the axis up to ± 30°

Leakage current ≤ ± 4 fA
Cable leakage ≤ 1 pC/(Gy·cm)

Materials and measures

Wall of sensitive volume 0.35 mm PMMA,
1.19 g/cm3
0.15 mm graphite,

0.82 g/cm3

Total wall area density 54 mg/cm2
Dimension of sensitive volume

radius 5 mm
length 18 mm

Central electrode Al 99, diameter 2 mm
Build-up cap PMMA, thickness 3 mm

Ion collection efficiency at nominal voltage

Ion collection time 40 µs
Max. dose rate for
≥ 99.5 % saturation
≥ 99.0 % saturation
54 Gy/s
105 Gy/s
Max. dose per pulse for
≥ 99.5 % saturation
≥ 99.0 % saturation
1.0 mGy
2.5 mGy

Useful Ranges

Chamber voltage ± (100 … 400) V
Radiation quality 70 keV … 50 MV photons
(10 … 45) MeV electrons
Field size (5 x 5) cm2 … (40 x 40) cm2

Temperature (10 … 40) °C
(50 … 104) °F
Humidity (20 … 75) %, max 20 g/m3

Air pressure (930 … 1050) hPa

Options

T48012 Radioactive check device 90Sr

T48002.3.004 Chamber holding device for check device

Parts

TN23332 Rigid stem chamber 0.3 cm3, connecting system BNT

TW23332 Rigid stem chamber 0.3 cm3, connecting system TNC

TM23332 Rigid stem chamber 0.3 cm3, connecting system M