0.1 Cm3 Semiflex C Chamber Type 23322

Therapy chamber for in-vivo measurements

0.1 Cm3 Semiflex C Chamber Type 23322
 

Features

  • Sealed ionization chamber for in-vivo measurements
  • Waterproof, flexible design for intracavitary use
  • Sensitive volume 0.1 cm3, fully guarded
  • Radioactive check device (option)

The 23322 C chamber is used for patient in-vivo measurements in conjunction with an electrometer classified for this purpose. Applications are external beam and brachytherapy dose measurements on the patient’s surface or in body cavities. Due to the small chamber diameter the chamber can be used for measurements in the bladder when used with a protective rubber sleeve. The C chamber is to be used with a radioactive check device and features a stable response even if the chamber temperature changes.

Specifications

Type of product cylindrical ionization chamber acc. IEC 60601-2-9
Application absolute dosimetry in radiotherapy beams
Measuring quantities absorbed dose to water, air kerma, exposure
Reference radiation quality 60Co
Nominal sensitive volume

0.1 cm3

Design waterproof, fully guarded

Reference point on chamber axis, 8.9 mm from chamber tip

Direction of incidence radial
Nominal response 3.8 nC/Gy
Long-term stability ≤ 2 % per year
Chamber voltage 400 V nominal
± 500 V maximal
Photon energy response < ± 4 % (280 kV … 60Co)
Directional response in water < ± 0.5 % for rotation around the chamber axis and for tilting of the axis up to ± 30° (280 kV)

Temperature dependence ≤ ± 5 % within the useful range
Temperature equilibrium time (2 … 3) min/°C
Leakage current ≤ ±4 fA
Cable leakage ≤ 1 pC/(Gy·cm)

Materials and measures

Wall of sensitive volume 0.59 mm PMMA,
1.19 g/cm3
0.13 mm graphite,

0.82 g/cm3
polyolefin

Total wall area density

122 mg/cm2

Dimension of sensitive volume radius 1.625 mm
length 11.875 mm
Central electrode Al 99.9, diameter 0.8 mm

Ion collection efficiency at nominal voltage

Ion collection time 28 µs
Max. dose rate for
≥ 99.5 % saturation
≥ 99.0 % saturation

107 Gy/s
213 Gy/s

Max. dose per pulse for
≥ 99.5 % saturation
≥ 99.0 % saturation
2.1  mGy

4.2  mGy

Useful Ranges

Chamber voltage ± (100 … 400) V
Radiation quality 66 keV … 50 MV photons
Field size (4 x 4) cm2 … (40 x 40) cm2

Temperature (10 … 40) °C
(50 … 104) °F
Humidity (20 … 75) %, max 20 g/m3

Air pressure (700 – 1060) hPa

Options

T48012 Radioactive check device 90Sr

T48002.1.006 Chamber holding device for check device

Parts

TN23322 semiflex C chamber 0.1 cm3, connecting system BNT

TW23322 semiflex C chamber 0.1 cm3, connecting system TNC

TM23322 semiflex C chamber 0.1 cm3, connecting system M

Image Gallery

Click on a thumbnail below to view full size.