Features
Cardio-C is an ultra-compact cardiac-optimized nuclear imaging system. It is comprised of two detectors in 90o geometry mounted inside a C-shaped gantry, a compact 4-axis imaging table and a dedicated acquisition station. Cardio-C is networked to Interview Processing & Review Station which features a comprehensive set of Cardiac applications
Clinical Applications
Cardio-C features a wide variety of cardiac imaging types, including static, dynamic, multigated planar, tomographic, and gated tomographic nuclear imaging procedures.
Specifications
SPECIFICATION PARAMETER DATA RANGE
- Intrinsic Energy Resolution UFOV FWHM < / = 9.7 %
- Intrinsic Spatial Resolution CFOV FWHM < / = 2.8 mm
- CFOV FWTM < / = 5.4 mm
- UFOV FWHM < / = 2.9 mm
- UFOV FWHM < / = 5.6 mm
- Intrinsic Flood Field Uniformity:
- Integral (Max. deviation) UFOV < / = ± 2.9 %
- CFOV < / = ± 2.4 %
- Differential (Max deviation) UFOV < / = ± 2.4 %
- CFOV < / = ± 1.9 %
- Intrinsic Spatial Linearity:
- Differential UFOV < / = 0.20 mm
- CFOV < / = 0.18 mm
- Absolute UFOV < / = 0.40 mm
- CFOV < / = 0.38 mm
- Intrinsic Count Ratev 20% count rate loss, 20% window
- Incident > / = 200 kcps
- Observed > / = 160 kcps
- Maximum C. Rate , 20% window
- > / = 220 kcps
- Multiple Window Spatial
- Registration (MWSR)
- < / = 1.5 mm
Product Number: Cardio-C