Low Energy Gamma Scintillation Probe

Low-Energy Gamma Scintillation Probe for Gamma and X-ray above 10 keV

Low Energy Gamma Scintillation Probe
 
  • Features
  • Specifications
  • Image Gallery

Features

Low-Energy Gamma Scintillation Probe for Gamma and X-ray above 10 keV

Specifications

Type:
NaI (Tl); Thin Scintillator for Low Energy Gamma, scintillator optically coupled to PMT

Radiation detected:

Gamma and x-ray above 10 keV

Applications:

  • Primary probe for nuclear medicine
  • Low energy x-ray manufacturing
  • Industrial hygiene

Typical background (CPM):
200

Nominal sensitivity:
3,000,000 CPM/µCi 129l

Wall material:
0.04 in Al, 1 mm thick

Window:
8 mg/cm2 Al

Sensitive area:
5 cm2

Crystal dim.:

1 in Ø

Probe diameter:
2 in

Probe length:
8.125 in

Cable length:
48 in

Operating voltage:
900 V

Calibration:
Sensitivity to 129I

Cal. Tolerance:
± 10%

Efficiency:
90 Sr 22%;
36Cl 8%;
241Am 8%;
133Ba 34%

Humidity range:
0 to 95%

Operating temp:
- 40° to + 120°F; – 40° to + 50°C Max. temp increase of 20°F/hr

Weight (approx.):
1.5 lb (0.68 kg)

Product Number: 425-110

Image Gallery

Click on a thumbnail below to view full size.

  • Default Image

Posted in Fluke | Tagged detector, nai, probe, scintillation, sodium-iodide

« Previous Next »
  • Manufacturers
  • Technologies

Featured Manufacturers

  • Biodex
  • Bionix
  • Capintec
  • Diacor
  • Exini
  • Fluke
  • International Specialty Products
  • IZI Medical Products
  • Mediso
  • PTW
  • Topex
  • Unfors
  • Vision RT
  • View all...

Technologies

  • Nuclear Medicine
  • Radiology
  • Radiation Therapy

Copyright © 2012 Harpell Associates Inc..

  • Home
  • About
  • News
  • Contact