Features
The very compact 245 simplifies the determination of actual kV for radiographic, fluoroscopic, and mammographic x-ray systems. This highly accurate meter can be used for a wide range of energies and will store up to ten readings. A remote LCD display makes fluoroscopic readings easy to obtain. With an optional BNC connector the meter can also be connected to an oscilloscope to monitor waveforms.
- Radiographic, Fluoroscopic, and Mammographic non-invasive measurement modes
- Dual methods of kVp measurements available in a single radiographic or mammographic exposure: (1) the average of kVpeaks, (2) effective kVp
- Use of patented Quadcell detector and autogain circuitry for unparalled accuracy in kVp measurements over a wide range of measurement conditions: orientation, positioning, angulation, and x-ray intensity have little effect on measurement results
- Measurements of kVp and time can be accurately performed over an x-ray intensity dynamic range of 5000:1 at any specific kV
- Internal correction factors for mammographic target/filter combinations
- Automatic resetting for each exposure
- Large LCD display
- Compact size that fits in your hand
Specifications
- Kilovoltage: Displayed Range: 22kV- 200 kV
- Calibrated Range: Radiographic: 50 – 140 kV; Mammographic: 25 – 35 kV
- Accuracy: Radiographic: +2% or 1 kV; Mammaographic: +(1kV + 2%)
- Reproducibility: Radiographic: + 0.5 kV; Mammographic: + 0.3 kV; Resolution: 0.1 kV
- Minimum Exposure Requirements: Radiographic: 25mA, 60 kVp
24 in.; Fluoroscopic: 3 mA, 80 kVp18 in.; Mammographic: 80 mA, 25 kVp @ 25 in.; Required Exposures: One; Display: 3 1¼2 digit LCD; Analog Output: Optional BNC connector (Special Order); Power: 9 V alkaline battery (included); Size: 2×4 x 7.5 in.; Weight: 21 oz.; Target/Filter Combinations: Mo/Mo, Mo/Ro, Mo/Al, Mo/Pd, W/Al, Ro/Ro, Ro/Al