Features
This Hot-Cell Monitor answers the need for multiple detectors when you want to locate and measure the activity. The unit monitors gamma and high energy beta emitters with activity levels from 1 mCi to 100 Curies. There is Remote operation for hot-cell and other high level radiation environments. Up to 8 detectors, with either linear or logarithmic output, are powered and monitored simultaneously. Front panel bar displays provide instant, visual information on the movement and accumulation of activity provided for computer interfacing.
The miniature, solid state detectors monitor activity at distances of rugged, easy to mount, easy to shield and connect to the base unit
- For PET, Chemistry, Nuclide Production and Purification
- Monitors activity in as many as 8 locations
- Reads activities from 1 mCi to 100 Curies
Specifications
Base Unit:
| Bar Display (1-8) | 10 LED display Log, 3 or 6 decade Linear, gain adjustable |
|---|---|
| Digital Meter | Detector assignable Log, 1.0 V / decade |
Back Panel:
| Detector Controls (1-8) | BNC connectors Log, zero adjust Linear, gain adjust |
|---|---|
| Analog Output (1-8) | 0 – 1V / decade Log, 1.0 V / decade Linear, gain adjustable BNC connectors |
Detector Standard:
| Type | PIN Diode / CsI/(Na) 1×1 cm2 effective area |
|---|---|
| Configuration | Metal housing, fully potted Dimensions, 2.5×2.5 cm dia. BNC connector |
| Sensitivity Range | 0.5 mCi – 100 Ci |
| Linearity | ±10%, 1 mCi – 5 Ci |
| Operating Temp. | 0 -60° C |
Detector Ultra-Miniature:
| Type | PIN Diode / CsI/(Na) 6×6 mm2 effective area |
|---|---|
| Configuration | Epoxy, fully potted Dimensions, 1.5×1.0 cm dia. Flying lead with BNC connector |
| Sensitivity Range | 1 mCi – 100 Ci |
| Linearity | ±10%, 5 mCi – 5 Ci |
| Operating Temp. | 0-50° C |
Parts
- 8400-0326 BNC cable, 2 meters
- 8400-0326 BNC cable, 2 meters
- 8400-0327 BNC cable, 4 meters
- 8400-0328 BNC cable, 6 meters
- 5250-0069 Hot-Cell Monitor Base Unit
- 5250-0063 Logrithmic channel amplifier electronics, 5 decade range, 1 mCi to 100 Ci
- 5250-0064 Linear channel amplifier electronics, 2.5 decade adjustable range
- 1200-0048 Standard, high sensitivity detector, full potted in aluminum housing, 2.5 cm x 2.5 cm (dia)
Image Gallery