Darklight Duo-Combination Sensitometer/Densitometer with automatic Scanning

Darklight® Duo scan with integrated PC interface, user menu, and automatic scanning

Darklight Duo-Combination Sensitometer/Densitometer with automatic Scanning
 

Features

  • For constancy tests of film processors according to German Standard DIN 6868 and International Standard IEC 61223-2-1
  • High precision and reliability
  • Sensitometer according to DIN 6868-2
  • Densitometer with digital display
  • RS232 Interface (9- poles serial socket) for data transmission to PC
  • Integrated user menu for guiding the user through the operation steps for constancy test
  • Stored reference values for up to 25 locations
  • Output of logarithmic speed and average gradient (LS and G) as defined in ISO 9236-1
  • Automatic function for connection measurements for new film batches
  • Output of new reference values according to DIN V 6868-55 (Dmin), sensitivity index SI, contrast ind. CI)
  • Automatical evaluation of all 21 sensitometer steps; measuring time less than 5 seconds!
  • Manual measurement at any place with 180 mm arm length

Specifications

Technical Data

  • Size: 240×165 x 60 mm
  • Weight: 800g (without batteries)
  • Power: 4 x AA/ 1.5 V (Alkaline or NiCd) or darklight ® power supply

Sensitometer Specifications

Specification by German Standard DIN 6868-2 and DIN V 6868-55

  • Spectral characteristic: blue/green switchable
  • Peak wave length: blue 460 nm + 10 nm; green 510 nm + nm
  • Exposure: adjustable in 5 steps, divided for every exposure color; +0.3 log exp.; +0.15 log exp.; -0.15 log exp; -0.3 log exp.
  • Long term exposure stability: + 0.02 log exp.
  • Grey wedge: 21 steps
  • Grey wedge constant: 0.15 log exp.
  • Homogenity of the steps: + 0.01 log exp
  • Precision of the steps: 5% wedge for duo scan; 1% wedge for duo scan ref; + 0.05 for D<1 (duo scan); + 0.01 for D<1 (duo scan ref); Step 1 covered (measuring value for vase and fog = Dmin)

Densitometer Specifications

Specification by German Standard DIN 6868-2, DIN V 6868-55 and International Standard IEC 61223-2-1

  • Spectral characteristics: visual optical density according to German standard DIN 4512-3; internal light source
  • Measuring orifice aperture: Ø 3mm
  • Measuring range: D = 0 up to D> 4.5
  • Measuring accuracy for D< 3: + 0.01 (max. + 0.02)
  • Repeat accuracy for D: + 0.01
  • RS 232 Interface (9-poles serial socket) for data transmission to PC
  • Integrated user menu
  • Measuring modes: manual single measurement; 21 step automatic passage; 21 step manual measurement
  • Film speed: manual on touch, typical 5 sec; speed variable 3-20 sec; all formats possible

Parts

  • 0695-0083 Darklight Sensitometer
  • 0695-0082 Darklight Densitometer
  • 0695-0084 Darklight Duo EC Combination Sensitometer and Densitometer
  • 0695-0085 Darklight Duo Combination Sensitometer and Densitometer with PC Interface
  • 0695-0086 Darklight Duo CombinationSensitometer and Densitometer with Automatic Scanning and PC Interface