Don’t worry about meeting the new regulations in 10 CFR Part 35.315 (200 dpm requirements for unrestricted areas and iodine contamination). Capintec’s CAPRAC Wipe-Test Counter handles all of them.
Measure for measure, no other well counter offers the speed, accuracy and complete range of built-in features provided by the compact CAPRAC. It performs a wipe test in just 6 seconds (for 1 nCi) and detects extremely low levels of activity with the accuracy only a NaI drilled-well detector can provide.
The CAPRAC can also serve as a single-well gamma counter in departments that do not need multi-sample changers. User-defined protocols, trigger levels, and counting times are a "snap". A 6-channel pulse-height analyzer permits built-in gamma spectroscopy. Definable conversion factors for specific radionuclides allow the CAPRAC to calculate results in cpm, dpm, nCi, cps, dps, or kBq. The CAPRAC also displays photon-energy histograms.
The CAPRAC is engineered and built for years of consistent, reliable performance and is Curie or Becquerel selectable. The unit has automatic background subtraction and self-diagnosis programs for systems testing with optimized signal-to-noise ratio. It includes a lead outer shield (1.3 cm thick) with optional auxiliary shield available.
- Meets NRC/Agreement State regulations, including new regulations in 10 CFR Part 35.315
- Simple to operate
- NaI drilled-well crystal detector
- Performs wipe test in 6 seconds
- 6-channel, pulse-height analyzer
- Preliminary isotope identification through gamma spectroscopy
- Accepts and stores individual trigger levels for different areas (hot lab, patient rooms, unrestricted, sealed-source leak tests, etc.)
- Optional printer for hard-copy archives wipe-test results; prints photon-energy histograms
- Counts which exceed trigger levels are printed in red
Specifications
- Counter: 16.5 lbs, 7.25" (18.4 cm) wide, 11.5" (29.2 cm) deep, 10.3" (26.2 cm) high
- NaI(Tl) crystal detector
- Power: Standard: 115VAC 50/60Hz 0.1A Optional: 220V 50/60Hz 0.05A
- Measurement periods: 6, 20, 60, 180, 600, 1800 sec
- Radiation shield: 0.5" (1.3 cm) lead outer shield
- Counting rate: cpm, kcpm (Ci) cps, kcps (Bq)
- Counting channels:
- 1 = 15-100 keV
- 2 = 100-200 keV
- 3 = 200-400 keV
- 4 = 400-660 keV
- 5 = 660-800 keV
- 6 = >800 keV
- Sources : Cs-137 or Ba-133 (optional)
- Maximum count rate 60,000cps
Parts
- 5430-2057 Caprac Wipe Test/Well Counter
- 0975-137R Cs-137 Rod Source 0.5µCi
- 0975-138S Ba-133 Rod Source 0.1µCi
Image Gallery
Click on a thumbnail below to view full size.